Recall of NovoPen Echo insulin cartridge

Pharmaceutical company Novo Nordisk has initiated a consumer level recall of some batches of their NovoPen Echo insulin cartridge holder after discovering there is a possibility they may crack or break if exposed to chemicals in certain cleaning agents.

A cracked or broken cartridge holder can result in the device delivering a smaller dose of insulin than the user expected, leading to high blood sugar levels and potentially putting the person at risk of hyperglycaemia.

Novo Nordisk has advised that if people clean the pens as described in the User Guide, there is no reason that cracking of the cartridge holder will occur.

The recall only relates to some batches of the cartridge holders and there are no reported problems with the insulin being administered. Novo Nordisk advises that the risk of patients experiencing high blood sugar levels due to an affected cartridge holder is extremely low.

Novo Nordisk is recalling all pens in the affected batches not currently issued to a patient. People with diabetes who use a NovoPen Echo should immediately check the batch number and if it is from one of the affected batches they should not use it but apply to Nova Nordisk for a replacement.